European Journal of Cardio-Thoracic Surgery, Vol 10, 1003-1009, Copyright © 1996 by European Association for Cardio-thoracic Surgery
S Nazari, F Luzzana, C Banfi, Z Mourad, S Salvi, A Gaspari and F Nazari-Coerezza
OBJECTIVE: Most complications of descending aorta prosthetic substitution
seem mainly to be related directly (ischemia to distal organs, i.e. liver,
kidney, spinal cord) or indirectly (extracorporeal circulation or shunts
and systemic heparinization complications) to the duration of blood flow
interruption. the purpose of this study is to report the results of animal
experimentation of a new device for sutureless prosthetic substitution of
the descending thoracic aorta, with a very short cross-clamping phase.
METHODS: The device consists of expandable loops of stainless steel wires,
sewn to the proximal end of a Dacron prosthesis. The stainless steel wire
loops can be expanded and tightened by activating a removable guide in such
a way that the prosthesis varies its diameter, while maintaining a regular
cylindrical shape. The device was prepared in two different configurations,
one for long segments (expandable prosthesis end) and the other to be used
for very short segments or as an anastomotic ring between prosthetic or
vascular stumps (quick anastomotic ring). The expandable prosthesis end was
tested in swine experiments by performing the prosthetic substitution of
the first 10 cm of descending cross-clamped aorta, the prosthesis being
fixed with the device both at the proximal and the distal ends (six
experiments). All animals survived the procedure, that was accomplished
with a very short cross-clamping time. The quick anastomotic ring was used
to anastomose two prosthesis ends, at the middle of the prosthetic segment
used for descending aorta substitution (two swine), to perform the distal
anastomosis in the same model of descending aorta substitution (one swine)
and simply to re-anastomose a subtotally transected descending aorta (one
swine). RESULTS: The present experience proved the reliability of the
device to carry out a sutureless, accurate, simple and quick anastomosis.
Its advantage over an intraluminal ringed prosthesis is much easier
insertion of the retracted wired end into the vascular stumps, thus
allowing for a prosthetic diameter appropriate to the substituted vessel.
CONCLUSIONS: The reduced cross-clamping feature of the device would suggest
its use mainly in thoracic aorta prosthetic substitution for the prevention
of ischemic damage to distal organs; it can also be used to advantage
wherever an end-to-end vascular or prosthetic anastomosis is indicated,
providing an accurate, stented anastomosis.
ARTICLES
Expandable prosthesis for sutureless anastomosis in thoracic aorta prosthetic substitution
Department of Surgery, IRCCS San Matteo, University of Pavia, Italy.
This article has been cited by other articles:
![]() |
S. Elkouri, P. Gloviczki, and N. W. Chbat Minimally Invasive Vascular Surgery and the Evolution of Vascular Anastomosis Techniques Perspectives in Vascular Surgery and Endovascular Therapy, January 1, 2002; 15(2): 127 - 153. [Abstract] [PDF] |
||||
![]() |
S. Nazari, S. Salvi, A. Aluffi, E. Visconti, G. Rescigno, and P. Buniva Prosthesis for Aortic Arch Substitution Ann. Thorac. Surg., November 1, 1997; 64(5): 1339 - 1344. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |