Objective: To determine the optimal venous tubing diameter for adult cardiopulmonary bypass (CPB) to improve gravity drainage and to reduce priming volume. Methods: (A) Maximum bovine blood flow rates by gravity drainage were assessed in vitro for four different tubing diameters (1/2, 3/8, 5/16,1/4 inch) with three different lengths and various pre- and afterloads. Based on the results of (A) and multiple regression analyses, we developed equations to predict tubing sizes as a function of target flows. (C) The equations obtained in (B) were validated by ex vivo bovine experiments. (D) The clinically required maximal flows were determined retrospectively by reviewing 119 perfusion records at Zurich University. (E) Based on our model (B), the clinical patient and hardware requirements, the optimal venous tubing diameter was calculated. (F) The optimized venous tubing was evaluated in a prospective clinical trial involving 312 patients in Hangzhou. Results: For a mean body surface area of 1.83±0.2 m2, the maximal perfusion flow rate (D) achieved with 1/2-inch (=1.27 cm2) venous tubing was 4.62±0.57 l/min (range: 2.506.24 l/min). Our validated model (B,C) predicted 1.0 cm2 as optimal cross-sectional area for the venous line. New tubing packs developed accordingly were used routinely thereafter. The maximal flow rate was 4.93±0.58 l/min (range: 3.97.0) in patients with a mean body surface area of 1.62±0.21 m2. Conclusion: The new venous tubing with 1.0-cm2 cross-sectional area improves the drainage in the vast majority of adult patients undergoing CPB and reduces the priming volume (-27 ml/m). Reduced hemodilution can prevent homologous transfusions if a predefined transfusion trigger level is not reached.
Key Words: Cardiopulmonary bypass Gravity drainage Venous line Priming volume Extracorporeal circulation
This article has been cited by other articles:
![]() |
L. K. von Segesser, G. Siniscalchi, K. Kang, O. Maunz, J. Horisberger, E. Ferrari, D. Delay, and P. Tozzi Temporary caval stenting improves venous drainage during cardiopulmonary bypass Interactive CardioVascular and Thoracic Surgery, December 1, 2008; 7(6): 1096 - 1100. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. von Segesser, E. Ferrari, D. Delay, O. Maunz, J. Horisberger, and P. Tozzi Routine use of self-expanding venous cannulas for cardiopulmonary bypass: benefits and pitfalls in 100 consecutive cases Eur. J. Cardiothorac. Surg., September 1, 2008; 34(3): 635 - 640. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Abdel-Sayed, J Favre, J Horisberger, S Taub, D Hayoz, and L. von Segesser New bench test for venous cannula performance assessment Perfusion, November 1, 2007; 22(6): 411 - 416. [Abstract] [PDF] |
||||
![]() |
A. F. Corno Systemic venous drainage: can we help Newton? Eur. J. Cardiothorac. Surg., June 1, 2007; 31(6): 1044 - 1051. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. von Segesser, P. Tozzi, E. Riccardo Ferrari, C. Huber, D. Delay, D. Jegger, and J. Horisberger Small access (30F) clinical central venous smart cannulation: is it adequate? Interactive CardioVascular and Thoracic Surgery, October 1, 2006; 5(5): 540 - 543. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K von Segesser, P. Tozzi, I. Mallbiabrrena, D. Jegger, J. Horisberger, and A. Corno Miniaturization in cardiopulmonary bypass Perfusion, July 1, 2003; 18(4): 219 - 224. [Abstract] [PDF] |
||||
![]() |
D. Jegger, A. F Corno, A. Mucciolo, G. Mucciolo, Y. Boone, J. Horisberger, I. Seigneul, M. Jachertz, and L. K von Segesser A prototype paediatric venous cannula with shape change in situ Perfusion, January 1, 2003; 18(1): 61 - 65. [Abstract] [PDF] |
||||
![]() |
X. M. Mueller, I. Mallabiabrena, G. Mucciolo, and L. K. von Segesser Optimized venous return with a self-expanding cannula: from computational fluid dynamics to clinical application Interactive CardioVascular and Thoracic Surgery, September 1, 2002; 1(1): 23 - 27. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. M. Mueller, D. Jegger, M. Augstburger, J. Horisberger, G. Godar, and L. K. von Segesser A new concept of integrated cardiopulmonary bypass circuit Eur. J. Cardiothorac. Surg., May 1, 2002; 21(5): 840 - 846. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |